DSC131 evo

The highlights

An-easy-to-use Differential Scanning Calorimeter DSC (-170 to 700°C) with a 3-year warranty

The HIGHLIGHTS
– Robustness: the DSC131 evo features a highly robust sensor that, if needed, can readily be changed by the user in less than 30 minutes. No other DSC offers such flexibility.

– Ease of operation: dedicated market leading CALISTO software that is not only intuitive but powerful enough to perform every typical experiment and data treatment.

– Flexible and powerful: from -170 to 700 °C, with rapid heating and cooling performance.

– Large range of crucibles: in addition to the regular 30 μl crucibles, crucibles with a capacity of 100 μl are used for analyzing heterogeneous samples and for optimizing Cp (heat capacity) measurements.

High Pressure Crucibles: The High Pressure crucibles deliver unmatched High Pressure resistance (up to 500 bars / 7 250 psi, 600 °C) while the DSC sensor itself remains at atmospheric pressure.

Specifications
Temperature rangeAmbient to 700 °C
With cooling accessories-170 °C to 500 °C (Liquid Nitrogen Accessory)
-70 °C to 400 °C (Cryothermostat)
Programmable temperature scanning rate(heating and cooling)0.01 to 100 °C.min-1
Cooling time12 min (500 °C to 100 °C) Air
5 min (100 °C to 0 °C) Cryothermostat
6 min (200 °C to 25 °C) Liquid Nitrogen Accessory
12 min (25 °C to -100 °C) Liquid Nitrogen Accessory
RMS Noise (200°C)1.5 μW
Resolution0.8 μW
Time constant3 s
Gasesincluded gas switch from gas A to gas B
Crucibles30 μl, 100 μl
aluminum, alumina, incoloy, etc.
Pressure (non controlled)High pressure crucible
(up to 500 bars / 7255 psi at 600 °C)
Weight37.4 kg (82.5 lbs)
Dimensions (Height / Width / Depth)40 / 53 / 58 cm (15.7 / 20.9 / 22.8 in)
Power requirements230 V – 50/60 Hz.
Features

CRUCIBLES
– Regular crucibles
We offer a range of crucibles designed to ensure good thermal transfer between the sample and the sensor – Alumina, Aluminum (30 and 100 μl).

– High pressure crucibles
Incoloy and Gold plated Incoloy (30 μl) for high pressure capability: the High Pressure Incoloy crucibles deliver unmatched High Pressure capability (up to 500 bars / 7 250 psi, 600 °C) while the DSC sensor itself remains at atmospheric pressure.

SUBAMBIENT COOLING SYSTEMS
For subambient temperatures, two types
of cooling device are available :

– A simple liquid nitrogen (LN2) cooling accessory
The LN2 accessory is a manually operated cooling accessory. Its temperature of operation is from -170 to 500 °C and it is a highly robust and cost effective solution to low temperature control.

– A cryothermostat cooling device for intermediate temperature ranges :
-80 °C to +200 °C under a flow of Helium -50 °C to +500 °C under a flow of Argon,Nitrogen or dry Air.

Applications

Applications
With its wide temperature range (-170 °C to 700 °C) the DSC131 evo can meet a wide range of applications, especially when dealing with polymers and plastics(characterization, quality control) as well as with organic and pharmaceutical compounds (polymorphism, purity, thermal stability), with inorganic substances (dehydratation, transition, decomposition), with metals (transition)… from the raw to transformed materials.

As a teaching tool the DSC131 evo offers a unique blend of performance, ease of use and robustness to allow the training of students in the thermodynamic principles of phase change(fusion, crystallization, evaporation), transition (glass transition, order-disorder transition), reaction kinetics(polymerization, decomposition), or heat capacity.

View the application notes in your field, available for download, by visiting our Application Library area

We have also included a powerful search engine that will enable you to find the most applicable data.

Documentation


Brochures
Posters
Application notes