| Temperature range(the use of an auxiliary thermostat may become necessary) |
-45°C / +120°C |
| Temperature accuracy |
+/-0.1°C |
| Temperature precision |
+/- 0.02°C |
| Programmable temperature scanning rate (heating and cooling) |
0.001 to 2°C.min-1 |
| Cooling time |
20 min (from 120°C to ambient) |
| Enthalpy accuracy |
+/-1% |
| Calorimetric precision |
+/-0.1% |
| RMS Noise |
0.4 µW |
| Sensitivity (Joule effect at 30°C) |
102 µV/mW |
| Resolution |
0.02 µW / 0.002 µW |
| Dynamic range |
+/-190 mW |
| Cells |
1 ml, made of Hastelloy C - Removable |
| Pressure (measured & controlled) |
requires the use of specific cells and gas panel 400 bar / 5800 psi |
| Weight |
37.4 kg (82.5 lbs) |
| Dimensions (Height / Width / Depth) |
40 / 53 / 58 cm (15.7 / 20.9 / 22.8 in) |
| Power requirements |
230V -50/60Hz |